Load Control
The use of the signal
from the loadcell in the control feedback loop permits the load/time history
of the sample to be controlled. In order to operate safely in this mode, it
is important to arrange for control to be transferred to POSITION CONTROL
(or the apparatus stopped) if the sample fails. Setting the allowed maximum
and minimum values of the load, and enabling an Event Detector to effect
this control transfer will ensure safe operation. If this is not done, the
servo will attempt to change its position to maintain the load at the requested
value. This will cause the hydraulic piston to hit the end of its range
and damage the machine.
WHEN USED IN LOAD CONTROL, THE EVENT
DETECTOR MUST BE SET TO TRANSFER CONTROL TO POSITION CONTROL
WHEN THE SAMPLE FAILS |
|